-40%

XLNT! Audioscan Verifit Real Ear Hearing Aid Analyzer w/ Current E3 Calibration

$ 2637.36

Availability: 14995 in stock
  • Model: Verifit 1
  • MPN: Does Not Apply
  • Type: Audiometer
  • All returns accepted: ReturnsNotAccepted
  • Condition: Excellent Condition with Current E3 Calibration!
  • Brand: Audioscan Verifit

    Description

    Being offered is this Excellent Condition Audioscan Verifit 1, Real Ear Hearing Aid Analyzer with accessories and Current e3 Diagnostics Calibration!  Please refer to the pictures for details as to what's included.
    Product description
    The Verifit is a hearing aid analyzer intended to be used by hearing care professionals such as audiologists and
    hearing aid specialists to verify the electro-acoustic performance of a hearing aid connected to a standard
    earphone coupler or while worn on the ear of the end user. It consists of:
    a) an acoustically-treated test chamber which houses a battery substitute (1.3 volt supply), 2 loudspeakers, a
    reference microphone for controlling the signal from the loudspeakers, a standardized earphone coupler
    connected to a measuring microphone for the purpose of measuring the sound level produced in the standard
    coupler by a hearing aid and a custom keypad.
    b) an analyzer/display unit housing a flat-panel video display, power supply, signal generation, measurement and
    control electronics and two sound-field loudspeakers.
    c) two real-ear measurement microphone assemblies, each housing a reference microphone for controlling the
    signal from the sound-field speakers and a probe microphone connected to a thin silicone tube which may be
    inserted into the ear canal for the purpose of measuring the sound level in the ear canal produced by a hearing
    aid.
    d) a miniature earphone used to measure the real-ear to coupler difference (RECD) useful in estimating the sound
    level produced in an individual ear from measurements in the standard earphone coupler.
    The analyzer attaches to the acoustic test chamber via two 15-conductor cables. The acoustic test chamber
    provides connectors for the microphones used for controlling the stimulus and measuring the response. For HIT
    mode, a reference microphone, coupler microphone and a battery simulator are normally connected via
    connectors provided inside the acoustic test chamber. For REM mode, two probe modules – each containing a
    reference microphone and a measurement microphone, and a real-ear to coupler difference (RECD) earphone are
    normally connected via connectors situated on the front of the acoustic test chamber. A connector and volume
    control for headphones, used to monitor coupler and probe microphone signals, are provided on the front right
    side of the test chamber.
    There are connectors on the rear of the main display unit for 4 USB devices (flash drives, a QWERTY keyboard,
    mouse, printer, barcode scanner), external speakers, an external HDMI monitor and connection to a network
    (LAN) (serial number 4671 and higher). Units with serial number below 4671 have connectors for 2 USB
    devices, an external VGA monitor, parallel printer, external speakers, a local area network (LAN) and an RS223
    serial port (not supported in this version of software).
    Electrical supply input requirement: 100 – 240 Vac 47 – 63 Hz 250VA
    A hospital grade grounded outlet is required.